Restoring the Initial Method Results Analyzing AMP ADP ATP UDP and GPT in Biological Extracts - Troubleshooting
April 2, 2012
/
/
/
/
/
Why Peak Shape May Deteriorate After a Few Days

If you notice that peak shape for AMP, ADP, ATP, UDP, and GPT in biological extracts worsens after a couple of days using a Cogent Diamond Hydride™ column, the issue may be related to sodium contamination in your LC-MS system.

Why This Happens:
Phosphate-containing metabolites are highly sensitive to sodium. Over time, sodium can leach from borosilicate glass mobile phase bottles, reaching levels that affect peak shape for this class of compounds.

Solutions:

  • Option A: Prepare fresh mobile phase daily.
  • Option B (Recommended): Switch to Teflon mobile phase reservoir bottles for a long-term fix.

If you have J.T. Baker Nitric Acid – ULTREX (Canada), it comes in ideal Teflon bottles. To prepare these bottles for LCMS use:

  1. Rinse with DI water (3 times).
  2. Fill with a 1:1 methanol/DI water solution and let sit overnight.
  3. After this, the bottles are ready for LC-MS.

Additional Tips:

  • Adjust aqueous ammonium acetate or formate mobile phase to pH 7.0 using ammonia.
  • Add 0.1% ammonia to the sample diluent to improve peak shape (note: sample stability may decrease).
  • Dedicate one column for ammonium acetate/formate and another for acidic mobile phases (e.g., formic acid).
Diamond Hydride Product Page
 

Related Articles

  1. Biological Extracts - Troubleshooting for Diminished Peak Height Using Cogent Diamond Hydride™ Columns - Tech Information
  2. Retention And or Peak Shape for Citric Acid has Changed Compared to Data from a Previous HPLC Run - Tips & Suggestions
  3. White Film on an LCMS Source with Background Noise was Very Low - Tech Information

© Copyright 2026. MICROSOLV. All Rights Reserved. Website & Hosting by BlueTone Media